MAIA Macular Integrity Assessment System from CenterVue

  • MAIA Macular Integrity Assessment System

Description

MAIA represents the latest advance in confocal microperimetry. Retinal images are acquired by Scanning Laser Ophthalmoscopy (SLO). An eye tracker allows accurate, real-time, compensation of eye movements. Luminance levels are compliant with existing standards (1000 asb.). MAIA is highly automated, very easy to use, non-mydriatic and combines the best aspects of its predecessors.

Similarly to Standard Automated Perimetry (SAP), MAIA measures light sensitivity of the retina by projecting Goldmann III stimuli over different retinal locations within the field of view. The main difference and great advantage over SAP is that retinal sensitivity is measured while simultaneously imaging the retina in real time. Retinal images, created by SLO technology, are processed by an eye tracker to calculate and compensate eye shifts (fixation losses) occurring during visual field measurement, in both physiological and pathological conditions. Fixation analysis is the second, fundamental, outcome of the MAIA eye tracker.

MAIA performs different types of microperimetry tests with supra and full-threshold strategies, and follow-up test to monitor functional progression. Each exam provides a measure of retinal sensitivity and fixation analysis (stability and position of the Preferred Retinal Locus). The MAIA sensitivity scale is 0 to 36 dB.
  • Product Specs
  • ItemMAIA Macular Integrity Assessment System
  • CompanyCenterVue
  • Price Inquire
  • Catalog NumberMAIA Macular Integrity Assessment
  • QuantityEA
  • Features
    • Simple to use, patients can be tested in less than 3 minutes per eye
    • By comparing macular indexes in patients with and without retinal pathologies, MAIA is very effective at measuring functional changes due to disease and to treatment
    • Helps monitoring the course of retinal diseases and the efficacy of treatment
    • Effectively combines a linear SLO system and a fundus controlled perimetric exam, providing clear and detailed retinal images
    • New Eye Tracker Alarm for the MAIA - The new MAIA software version 1.5.3 includes the innovative “Eye Tracker Alarm,” a feature that advises the operator when the patient is not properly following the MAIA test. This feature reduces even more the operator’s involvement during screening
  • InterfaceTouch screen
  • Weight (lbs)50.7 lbs (23 kg)
  • Resolution1024 x 1024 pixel
  • Working Distance33 mm
  • Field of ViewLine scanning laser ophthalmoscope: 36° x 36°
    For macular perimetry: 20° x 20°
  • Image Acquisition Speed25 fps
  • FocusAuto-focus (-15D to +10D)
  • Electrical Requirements100-240 VAC, 50-60 Hz, fuse 3.15 A (T type)
  • Power Consumption300 VA
  • Dimensions348 x 580 x 600 mm (13.7 x 22.8 x 23.6 in)